No relevant resource is found in the selected language.
Your browser version is too early. Some functions of the website may be unavailable. To obtain better user experience, upgrade the browser to the latest version.
Huawei uses machine translation combined with human proofreading to translate this document to different languages in order to help you better understand the content of this document.
Note: Even the most advanced machine translation cannot match the quality of professional translators.
Huawei shall not bear any responsibility for translation accuracy and it is recommended that you refer to the English document (a link for which has been provided).
Vacuumizing
Vacuumizing
Prerequisites
Before vacuumizing, check that the refrigerant pipe system has passed the air tightness test and does not leak.
Before vacuumizing, ensure that all connectors on the device are tightened.
System high pressure may occur if the system is not vacuumed or is not thoroughly vacuumed. Therefore, ensure that the system is thoroughly vacuumed.
Procedure
Check whether a low temperature component is installed on the outdoor side.
If no low temperature component is installed on the outdoor side, connect the pressure gauge and vacuum pump at the low-pressure and discharge pipe needle valves to vacuumize the system simultaneously, as shown in Figure 6-1.
Figure 6-1 Vacuumizing the system without a low temperature component
(1) Vacuum pump
(2) Pressure gauge
(3) Low-pressure needle valve
(4) Discharge pipe needle valve
If a low-temperature component is installed on the outdoor side, connect the pressure gauge and vacuum pump at the low-pressure, exhaust pipe, and liquid pipe needle valves to vacuumize the system simultaneously, as shown in Figure 6-2.
Figure 6-2 Vacuumizing the system with a low temperature component
(1) Vacuum pump
(2) Pressure gauge 1
(3) Pressure gauge 2
(4) Low-pressure needle valve
(5) Discharge pipe needle valve
(6) Liquid pipe needle valve
In the beginning, the vacuum pump makes loud noises and exhausts white gas from the exhaust vent. After 10 minutes, if it still exhausts white gas, observe it for another 10 minutes because the cooling system may not be sealed properly or there may be too much residual refrigerant or water in the cooling system.
After 20 minutes, the pressure gauge pointer should be in the negative area and the vacuum pump makes small noise. Close and open the pressure gauge alternatively for several times. The position of the pressure gauge pointer and the sound made by the vacuum pump do not obviously change. Otherwise, the cooling system may not be sealed properly.
Vacuumize the cooling system for at least 80 minutes after checking that the cooling system does not leak. The pressure displayed on the vacuum pump should be less than or equal to 60 Pa (absolute pressure). When the pressure stops dropping, vacuumize for another 10 minutes, after which the sight glass should display green, a sign of dryness.
After vacuumizing is complete, close all valves of the pressure gauge and the vacuum pump without disconnecting the pump, and preserve the pressure for 10 minutes. Ensure that the pressure of the cooling system does not exceed 90 Pa (absolute pressure).
If the minimum reading of the pressure gauge is greater than 60 Pa (absolute pressure), ensure that the pointer stays at the smallest scale of the pressure gauge in vacuumizing and preserve the pressure for 1 hour. Then check that the pressure does not rise evidently.